IMPORTANT: This supplement is intended as a convenient reminder of limited product information only. For complete product and maintenance information, see instruction booklet.
KEEP HEAD OF THE HANDPIECE UP IN VERTICAL POSITION
Scrub handpiece with soft brush and 70% alcohol to remove debris
OBSERVE ALL CAUTIONS LISTED
Towel dry handpiece thoroughly
Using a well-shaken can of handpiece conditioner with lube nozzle, attach handpiece and apply conditioner for two seconds over a towel or sink. With bur in place, run handpiece at full speed without water for 45 seconds to expel excess conditioner. Dry exterior of handpiece with a towel.
Remove bur, bag handpieces individually and autoclave or chemiclave per manufacturer’s instructions.
DO NOT exceed 275°F (135°C)
After the sterilization cycle please remove pieces when they’re completely dry to cool off. Do not use any handpiece while it’s HOT or WARM.
Caution: Handpiece may be too hot to handle!
When handpiece is cool to the touch, gently clean fiber optic light transmiting surfaces on both ends of handpiece with cotton swab dampened with isopropyl alcohol.
Remove attachment from air motor. Remove heads from contra angle/straight attachments. Remove all burs, prophy cups, etc. from heads and straight attachments.
KEEP HEAD OF THE HANDPIECE UP IN VERTICAL POSITION
Scrub handpiece with soft brush and 70% alcohol to remove debris
OBSERVE ALL CAUTIONS LISTED
Towel dry handpiece thoroughly
Apply handpiece conditioner to all lowspeed heads, attachments, and motors. Follow specific instructions detailed in instruction booklet. It is not necessary to lube motors or straight handpieces marked “No Lube Required.”
Insert all components into autoclave bag and load into autoclave/chemiclave.
DO NOT exceed 275°F (135°C)
After the sterilization cycle please remove pieces when they’re completely dry to cool off. Do not use any handpiece while it’s HOT or WARM.
To maintain your handpieces, we recommend the use of MHR Lubricant Drops instead of spray.